信息化融合六西格玛提升血常规检测室内质控效能

曹依蔚 ,吴梓文 *
作者信息
  1.  复旦大学附属儿科医院  上海  201102

摘要

目的:研究信息化融合六西格玛管理对常规检测室内质控效能的提升价值。方法:选取2024年6月至2025年12月复旦大学附属儿科医院检验科白细胞(white blood cell,WBC)、红细胞(red blood cell,RBC)、血红蛋白(hemoglobin,HGB)、血细胞比容(hematocrit,HCT)、血小板(platelet,PLT)、平均红细胞体积(mean corpuscular volume,MCV)、平均红细胞血红蛋白量(mean corpuscular hemoglobin,MCH)、平均红细胞血红蛋白浓度(mean corpuscular hemoglobin concentration,MCHC)共8项参数进行分析,在低、中、高3个浓度水平采集室内质控数据,应用六西格玛质量评价模型融合信息化技术进行质控分析和评价。结果:8项参数靶值符合率均>90%。WBC、MCV三水平均达世界一流6σ水平;HCT、MCH为优秀水平(5≤σ<6);RBC、MCHC为良好水平(4≤σ<5);低值HGB和低值PLT为临界水平(3≤σ<4)。所有项目均达临床可接受水平(σ≥3),质控修正率与失控排查率均达89%~95%。结论:信息化融合六西格玛可有效精确匹配质控规则,对提升血常规检测室内质控效能、保证检测质量有重要意义。

Informatization Integrated with Six Sigma for Improving the Efficacy of Internal Quality Control in Blood Routine Testing

Yiwei Cao , Ziwen Wu *
Author Information
  1.  Children’s Hospital of Fudan University, Shanghai 201102, China

Abstract

Objective: To study the value of informatization integrated with Six Sigma management in improving the efficacy of internal quality control for routine testing. Methods: A total of 8 parameters including white blood cell (WBC), red blood cell (RBC), hemoglobin (HGB), hematocrit (HCT), platelet (PLT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC) in the laboratory department of Children’s Hospital of Fudan University, from June 2024 to December 2025 were selected for analysis. Internal quality control data were collected at low, medium, and high concentration levels, and the Six Sigma quality evaluation model integrated with informatization technology was applied for quality control analysis and evaluation. Results: The target value compliance rates of all 8 parameters were >90%. WBC and MCV reached the world-class 6σ level at all three levels; HCT and MCH were at the excellent level (5≤σ<6); RBC and MCHC were at the good level (4≤σ<5); low-value HGB and low-value PLT were at the critical level (3≤σ<4). All items reached the clinically acceptable level (σ≥3), and both the quality control correction rate and the out-of-control investigation rate reached 89%–95%. Conclusion: Informatization integrated with Six Sigma can effectively and precisely match quality control rules, and is of great significance for improving the efficacy of internal quality control in blood routine testing and ensuring testing quality.